Labcyte Echo 520 User Manual page 151

Liquid handler
Table of Contents

Advertisement

Caution: When the F
strongly recommends that the user execute the Focus calibration
wizard as soon as possible.
1. In the T
RANSDUCER FOCUS CALIBRATION
Figure 5.47 Transducer focus calibration page 1
2. Accept all of the default values and do not select the O
C
R
ALCULATED
ANGE
3. Click the N
button.
EXT
The Echo process door opens, the source plate gripper stage extends,
and the Echo liquid handler performs a focal-sweep. The focal-sweep
uses acoustic "pings" bounced off an internal registration plate while
the transducer is being moved vertically. The focal-sweep process
measures the transducer's focal length. After the focal-sweep is done,
the bottom of the registration plate is dried.
4. Click the F
button to complete the focus calibration procedure.
INISH
C
OCUS
ALIBRATION
window, click the F
checkbox.
indicator is yellow, Labcyte
OCUS
VERRIDE
Echo Liquid Handler Software 5-51
button.