AMI Phosphate-II
Installation
3.2.
3.3.
Sample inlet
Mounting of
SERTO fitting
Waste
18
Mounting of Instrument Panel
Mount the instrument in vertical position. For ease of operation
mount it so that the display is at eye-level. For dimensions see pic-
ture
Dimensions Phosphate-II, S. 13
Connecting Sample and Waste
Use plastic tube (FEP, PA, or PE 4 x 6 mm) to connect the sample
line. See also
, S. 30
A
B
C
D
Connect the 1/2" tubes to the nozzle of the waste funnels and place
them into an atmospheric drain of sufficient capacity.
A
B
WARNING
the sample which flows after the measurement through the out-
let of the photometer contains Hexammonium heptamolybdate
4-hydrate.
At no means recirculate it into the water system.
A
Screw connection
B
Compression ferrule
C
Knurled nut
D
Flexible tube
A
Tube from photometer
B
Waste (photometer)
C
C
Tube from constant head
D
Drain (constant head)
D
A-96.250.641 / 290312