Mounting of Instrument Panel

Mount the instrument in vertical position. For ease of operation
mount it so that the display is at eye-level. For dimensions see pic-
ture
Dimensions Phosphate-II, S. 13

Connecting Sample and Waste

Use plastic tube (FEP, PA, or PE 4 x 6 mm) to connect the sample
line. See also
, S. 30
A
B
C
D
Connect the 1/2" tubes to the nozzle of the waste funnels and place
them into an atmospheric drain of sufficient capacity.
A
B
WARNING
the sample which flows after the measurement through the out-
let of the photometer contains Hexammonium heptamolybdate
4-hydrate.
At no means recirculate it into the water system.
A
Screw connection
B
Compression ferrule
C
Knurled nut
D
Flexible tube
A
Tube from photometer
B
Waste (photometer)
C
C
Tube from constant head
D
Drain (constant head)
D
A-96.250.641 / 290312