Replacing Fuses - Swann AMI Phosphate-II Operator's Manual

Version 5.01 and higher
Hide thumbs Also See for AMI Phosphate-II:
Table of Contents

Advertisement

AMI Phosphate-II
Maintenance

6.11. Replacing Fuses

A-96.250.641 / 290312
WARNING
Beware of electrical shock.
Disconnect the instrument and devices connected to:
 Relay 1
 Relay 2
 Alarm relay
from the mains before opening the cover.
Find and repair the cause for the short circuit before replacing the
fuse.
Use tweezers or needle-nosed pliers to remove the defective fuse.
Use original fuses provided by SWAN only.
A
B C D
A
1 AT/250V Instrument power supply
B
1 AT/250V Relay 1
C
1 AT/250V Relay 2
D
1 AT/250V Alarm relay
E
1 AF/125V Signal output 2
F
1 AF/125V Signal output 1
E
F
55