Swann AMI Phosphate-II Operator's Manual page 61

Version 5.01 and higher
Hide thumbs Also See for AMI Phosphate-II:
Table of Contents

Advertisement

AMI Phosphate-II
Error List
Error
E017
Control Timeout
E018
Reagent Pump
E019
FOME not connected
E020
FOME dirty
E021
Meas. invalid
E022
Reagent empty
E023
Cleaning Solution
E024
Input active
E025
IC MK41T56
E026
IC LM75
E027
IC PCF8574
E028
EEProm Microcon
E029
EEProm Motherboard
E030
EEprom Frontend
E031
Calibration Recout
E032
Wrong Frontend
E033
Sample Flow 1 low
A-96.250.641 / 290312
Description
Corrective action
– check control device or programming in
Installation, Relay contact, Relay 1/2
5.3.2/3, S. 77
– shut off power
– check wiring
– shut off power
– check wiring
– clean photometer, see
Photometer, S. 46
– this error disappears after the first
measuring
– refill reagents, see
Reagents, S. 41
– refill Cleaning Solution
– See If Fault Yes is programmed in
Menu
5.3.4, S. 79
– Hardware failure, call service
– Hardware failure, call service
– Hardware failure, call service
– Hardware failure, call service
– Hardware failure, call service
– Hardware failure, call service
– Hardware failure, call service
– Hardware failure, call service
nd
– Check 2
sample stream option
Cleaning the
Refill / replace
59