Swann AMI Phosphate-II Operator's Manual page 70

Version 5.01 and higher
Hide thumbs Also See for AMI Phosphate-II:
Table of Contents

Advertisement

AMI Phosphate-II
Program List and Explanations
2.2.1.5
2.2.1.5.1
2.2.2
2.2.2.1
2.3 Sample
2.3.1
2.4 I/O State
2.4.1/2.4.2
2.5 Interface
68
Ver. History: Review verifications values of the last verifications:
Number: Verification counter
Date, Time: Date and time of verification
Absorbance: Measured absorbance of the reference kit.
Reference value: True value of the reference kit according to label.
Miscellaneous:
Case Temp: Shows actual temperature in [°C] inside the transmit-
ter.
State Machine: Each number is assigned to a step of the measur-
ing cycle.
Sample ID: Shows the assigned sample identification. The identifi-
cation is defined by the user to identify the sample point in the
plant.
Sample Flow: Shows the actual sample flow in [B/s] (bubbles per
second). Sample flow must be above 5 B/s.
Raw value: Shows the raw value of the sample flow in Hz.
Shows actual status of all in- and outputs.
Alarm Relay:
Relay 1/2:
Input:
Signal Output 1/2:
Signal Output 3:
Only available if optional interface is installed.
Review programmed communication settings.
Open or closed.
Open or closed.
Open or closed.
Actual current in [mA]
Actual current in [mA] (if option is installed)
A-96.250.641 / 290312