AMI Phosphate-II
Installation
Sample
Connection
Electrical
Connection
A-96.250.641 / 290312
1 Stop sample flow.
2 Switch off the instrument.
3 Empty the flow cell.
4 Remove the sample inlet tube [E] from the flow cell block [A].
nd
5 Screw the 2
sample stream option (G) with the two fixing
screws (I) to the panel.
6 Install the tube [D] between the 2
the flow cell block.
7 Connect sample inlet 1 (J) and sample inlet 2 (K) to the
corresponding inlets at the 2
8 Connect the overflow tube (L) to the drain.
WARNING
Electrical shock hazard!
Before opening the AMI Transmitter switch power off.
Use one of the PG7 cable glands to feed the cable of the solenoid
valve into the AMI transmitter housing.
1 Remove the plug [A] from the cable gland [B.
2 Feed the cable of the solenoid valve through the cable gland (B)
into the AMI transmitter housing.
3 Connect the wires to the terminals in the AMI transmitter
according to the
nd
nd
sample stream option.
A
B
Connection Diagram Phosphate-II B, S.
sample stream option and
22.
25