Tube Replacement; Changing Pump Tubes - Swann AMI Phosphate-II Operator's Manual

Version 5.01 and higher
Hide thumbs Also See for AMI Phosphate-II:
Table of Contents

Advertisement

AMI Phosphate-II
Maintenance

6.10. Tube Replacement

6.10.1 Changing Pump Tubes

52
The pump tubes [D] of the peristaltic pump are exposed to a mini-
mal wear. It is therefore recommended to exchange the pump
tubes annually.
The long tubes, connected to the left (suction) side, leads to the re-
agent container, the short tubes, connected to the right (delivery)
side, leads to the flow cell.
ATTENTION
Change pump tube once a year.
WARNING
Never let the pump run during any manipulation, e.g replacing
reagent tubes, removing or mounting of the occlusion frame.
A
Pump housing
B
Occlusion frames
A
closed
C
Rotor
B
D
Pump tubes
C
E
Pump inlet
D
F
Pump outlet
E
F
Tube Numbering, S. 54
A-96.250.641 / 290312