Program List And Explanations; Messages; Diagnostics - Swann AMI Phosphate-II Operator's Manual

Version 5.01 and higher
Hide thumbs Also See for AMI Phosphate-II:
Table of Contents

Advertisement

AMI Phosphate-II

Program List and Explanations

9.

1 Messages

1.1 Pending Errors
1.1.5
1.2 Maintenance List
1.2.5
1.3 Message List
1.3.1

2 Diagnostics

2.1 Identification
2.1.3
2.1.4
2.1.5
2.2 Sensors
2.2.1
2.2.1.4
2.2.1.4.1
A-96.250.641 / 290312
Program List and Explanations
Provides the list of active errors with their status (active, acknowl-
edged). If an active error is acknowledged, the alarm relay opens
again. Cleared errors are moved to the Message list.
Demands necessary maintenance, e.g. preparing new reagents.
Shows the error history: Error code, date / time of issue and status
(active, acknowledged, cleared). 65 errors are memorized. Then
the oldest error is cleared to save the newest error (circular buffer).
In diagnostics mode, the values can only be viewed, not modified.
Designation: View the Designation of instrument.
Version: Firmware of instrument (e.g. V5.01 – 03/12)
Peripherals: PeriClip: Firmware of peristaltic pump (e.g. 1.03)
Factory Test: Test date of the Instrument -, Motherboard - and
Frontend - QC factory test.
Operating Time: Years / Days / Hours / Minutes / Seconds
FOME Sensor:
Current value: Shows the actual photometer signal [ppm].
Raw value: Shows the actual photometer signal [Hz].
Absorbance: Process value, depends on sample.
Cal. History: Review diagnostic values of the last calibrations.
Number: Calibration counter
Date, Time: Date and time of calibration
Slope: Slope photometer: 0.8–1.2