AMI Phosphate-II
Error List
7.
A-96.250.641 / 290312
Error List
Error
Non-fatal Error. Indicates an alarm if a programmed value is ex-
ceeded.
Such Errors are marked E0xx (bold and black).
Fatal Error
(blinking symbol)
Control of dosing devices is interrupted.
The indicated measured values are possibly incorrect.
Fatal Errors are divided in the following two categories:
Errors which disappear if correct measuring conditions are
recovered (i.e. Sample Flow low).
Such Errors are marked
Errors which indicate a hardware failure of the instrument.
Such Errors are marked
HOLD
R1
0.42 ppm
R2
15 B/s
Enter
Main Menu
Messages
Diagnostics
Maintenance
Enter
Operation
Installation
Messages
Pending Errors
Maintenance List
Message List
Enter
Pending Errors
Error Code
Sample Flow low
<Enter> to Acknowledge
E0xx
(bold and orange)
E0xx (bold and
Error
14:10:45
Error not acknowledged yet.
Check Pending Errors 1.1.5,
take corrective action.
25.4°C
(blinking) Fatal Error
Fatal Error not acknowledged yet.
Check Pending Errors 1.1.5,
1
take corrective action.
Reagent level low
Indicates the remaining reagent in
percent.
1.1
1.1.5
E010
red)
57