6.5.
44

Calibration

Select Menu 3.1 (Maintenance/Calibration) and follow the instruc-
tions on the display.
Calibration
Close sample tap
- prepare 1 liter of standard.
Standard: 1.00 ppm
Wait for next prompt
Press <Enter> to save the value or leave the menu with <Exit>.
Calibration
3.1.1
- Fill flow cell with
standard
Enter
<Enter> to continue
Calibration
Wait till calibration is
finished
Progress
Calibration
Calibration completed
Wait till flow cell is empty
Wait for next prompt
Calibration
Operation completed
Open sample tap
Slope
<Enter> to save
3.1.1
3.1.1
3.1.1
3.1.1
1.006
A-96.250.641 / 290312