AMI Phosphate-II
Installation
3.12. Signal Outputs
3.12.1 Signal Output 1 and 2 (current outputs)
3.12.2 Signal Output 3 (optional)
A-96.250.641 / 290312
Note: Max. burden 510
If signals are sent to two different receivers, use signal isolator
(loop isolator).
Signal output 1: Terminals 14 (+) and 13 (-)
Signal output 2: Terminals 15 (+) and 13 (-)
For programming see
Requires the additional board for the third signal output 0/4–20 mA
PCB
Note: Max. burden 510
The third signal output is installed in the upper holder on the main
board. You can operate either 3. signal output or communication in-
terface, not both!
Terminal 38 (+) and 37 (-).
.
A Third signal output 0/4 - 20 mA PCB installed on main board
Program Overview, S.
.
A
54, Menu Installation
29