Swann AMI Phosphate-II Operator's Manual page 50

Version 5.01 and higher
Hide thumbs Also See for AMI Phosphate-II:
Table of Contents

Advertisement

AMI Phosphate-II
Maintenance
Cleaning
48
WARNING
Fragile; handle with care.
1 Switch off the instrument according to instructions in
Operation for Maintenance, S. 40
2 Remove the constant head cover [A].
3 Pull the overflow tube [B] out of the flow cell block [D]
4 Remove the outer tube [C] from the flow cell block.
5 Clean all acrylic parts with a soft brush (bottle cleaner) and
soapy water.
6 Remove lime deposits with a common household deliming
agent with standard concentrations.
Note: A film of teflon paste (e.g. Fomblin from Solvay Solexis)
on the gaskets improves tightness and life time.