AMI Phosphate-II
Error List
Error
E001
Phos. 1 Alarm high
E002
Phos. 1 Alarm low
E003
Phos. 2 Alarm high
E004
Phos. 2 Alarm low
E005
Absorbance too high
E009
Sample Flow high
E010
Sample Flow low
E013
Case Temp. high
E014
Case Temp. low
E015
Valve defective
58
Description
Corrective action
– check process
– check programmed value
S. 76
– check process
– check programmed value
S. 76
– check process
– check programmed value
S. 76
– check process
– check programmed value
S. 76
– Clean Photometer
Photometer, S. 46
– check sample flow
– check programmed value
76
– adjust sample flow with the flow
regulating valve
– establish sample flow
– clean instrument
– check programmed value
76
– check environment temperature
– check programmed value
– check environment temperature
– check programmed value
– check valve, see 
5.3.1.1.x.1,
5.3.1.1.x.2x,
5.3.1.1.x.1,
5.3.1.1.x.2x,
Cleaning the
5.3.1.2.2, S.
5.3.1.2.3x, S.
5.3.1.4, S. 76
5.3.1.4, S. 76
50
A-96.250.641 / 290312