Input; Relay Contacts; Alarm Relay; Relay Contacts 1 And 2 - Swann AMI Phosphate-II Operator's Manual

Version 5.01 and higher
Hide thumbs Also See for AMI Phosphate-II:
Table of Contents

Advertisement

AMI Phosphate-II
Installation

3.10. Input

3.11. Relay Contacts

3.11.1 Alarm Relay

3.11.2 Relay Contacts 1 and 2

A-96.250.641 / 290312
Note: Use only potential-free (dry) contacts.
Terminals 30/31 at AMI Phosphate-II transmitter
Terminals 9/10 at AMI Phosphate-II B peristaltic pump
If signal output is set to hold, measurement is interrupted if input is
active.
For programming see menu
Programming of the relay contacts see
Note: Max. load 1 AT / 250 VAC
Alarm output for system errors.
Error codes see
Error List, S. 60
Programming see menu
Terminals
Normally
10 / 11
a)
open
Normally
12 / 11
closed
a) usual use
Note: Rated load 1 AT / 250 VAC
Relay 1: Terminals 6/7
Relay 2: Terminals 8/9
Programming see menu
5.3.4, S. 82
5.3 Relay Contacts, S. 78
5.3.1, S. 78
Description
Opened during normal operation.
Closed on error and loss of power.
Closed during normal operation.
Opened on error and loss of power.
5.3.2/3, S. 80
27