Cleaning The Solenoid Valve - Swann AMI Phosphate-II Operator's Manual

Version 5.01 and higher
Hide thumbs Also See for AMI Phosphate-II:
Table of Contents

Advertisement

AMI Phosphate-II
Maintenance
6.9.
Disassemble
the solenoid
valve
50

Cleaning the solenoid valve

The solenoid valve is mounted below the constant head. The sole-
noid valve should be disassembled if it does not switch anymore or
if it is clogged.
1 Switch off the instrument according to instructions in
Operation for Maintenance, S. 40
2 Loosen the nut (A).
A
3 Remove the solenoid coil (B) from
the valve body (C).
B
C
4 Loosen the fixing screws of the
valve body with a 2.5 mm Allen key
(D).
D
Stop of
A-96.250.641 / 290312