Operation - Swann AMI Phosphate-II Operator's Manual

Version 5.01 and higher
Hide thumbs Also See for AMI Phosphate-II:
Table of Contents

Advertisement

AMI Phosphate-II
Program List and Explanations

4 Operation

4.1 Sensors
4.1.1
4.1.2
4.2 Relay Contacts
4.3 Logger
4.3.1
Interval
Time
4.3.2
A-96.250.641 / 290312
Filter Time Constant: Used to damp noisy signals. The higher the
filter time constant, the slower the system reacts to changes of the
measured value.
Range: 5–300 sec
Hold after Cal: Delay permitting the instrument to stabilize again af-
ter calibration. During calibration- plus hold-time, the signal outputs
are frozen (held on last valid value), alarm values, limits are not ac-
tive.
Range: 0–6'000 sec
See
5.3 Relay Contacts, S. 75
The instrument is equipped with an internal logger. The data can be
downloaded to a PC by Hyperterminal if option "HyperTerminal in-
terface" is installed.
The logger can save approx. 1500 data records. Records consists
of: Date, time, alarms, measured value, measured value uncom-
pensated, temperature, flow.
Range: 1 second to 1 hour
Log Interval: Select a convenient log interval. Consult the table be-
low to estimate the max logging time. When the loggin buffer is full,
the oldest data record is erased to make room for the newest one.
(circular buffer)
1 s
5 s
25 min
2 h
Clear Logger: If confirmed with yes, the complete logger data is de-
leted. A new data series is started.
1 min
5 min
25 h
5 d
10
30
min
min
10 d
31 d
1 h
62 d
71