Output/Control - Swann AMI Phosphate-II Operator's Manual

Version 5.01 and higher
Hide thumbs Also See for AMI Phosphate-II:
Table of Contents

Advertisement

AMI Phosphate-II
Program List and Explanations
5.3.4
5.3.4.1
5.3.4.2
5.3.4.3
5.3.4.4
5.3.4.5
A-96.250.641 / 290312
Input: The functions of the relays and signal outputs can be de-
fined depending on the position of the input contact, i.e. no function,
closed or open.
Active: Define when the input should be active:
No:
Input is never active.
When closed:
Input is active if the input relay is closed
When open:
Input is active if the input relay is open
Signal Outputs: Select the operation mode of the signal outputs
when the relay is active:
Continuous:
Signal outputs continue to issue the measured
value.
Hold:
Signal outputs issue the last valid measured
value.
Measurement is interrupted. Errors, except fatal
errors, are not issued
Off:
Set to 0 or 4 [mA] respectively. Errors, except
fatal errors, are not issued.
Output/Control: (relay or signal output):
Continuous:Controller continues normally.
Hold:
Controller continues on the last valid value.
Off:
Controller is switched off.
Fault:
No:
No message is issued and the alarm relay does
not close when input is active.
Yes:
Message E024 is issued and alarm relay closes
when input is active.
Delay: Time which the instrument waits, after the input is deactivat-
ed, before returning to normal operation.
Range: 0–6'000 sec
79