Advertisement

AMI Phosphate-II
Program List and Explanations

5 Installation

5.1 Sensors
5.1.1
5.1.2
5.1.3
5.1.4
5.1.5
5.1.6
5.1.7

5.2 Signal Outputs

5.2.1 - 5.2.2
5.2.1.1-5.2.2.1
5.2.1.2-5.2.2.2
72
Ref. Verification: Set absorbance value of verification kit according
to label.
Range: 0.200–0.600
Phosphate as: Default setting is PO4. If the setting is changed from
PO4 to P, the measured value is expressed in P.
Available values: P, PO4
Standard PO4: Usual concentration range:
Range: 0.10 ppm to 8.00 ppm
Meas. Interval: Set the measuring interval:
Range: 10 min to 30 min
Channels: for AMI Phosphate-II B only. If the 2
option is installed, you can choose 1 or 2 channels.
Channel Selection: The following 3 operating modes can be set:
 Internal
 External
 Fieldbus
Mode Internal: In this mode, the AMI Phosphate-II B sequentially
measures each sample stream of the second stream option.
Mode External: In this mode, the AMI Phosphate-II B can be con-
trolled via signal input.
Mode Fieldbus: The AMI Phosphate-II B is controlled via fieldbus
Cleaning: Only available for AMI Phosphate-II and only visible if a
cleaning module is connected.
Program if one or two solutions are used on cleaning module.
Range: 1 Solution or 2 Solutions
Signal Output 1/2: Assign process value, the current loop range
and a function to each signal output.
Parameter: Assign one of the process values to the signal output.
Available values: Phosphate, Sample Flow.
Current Loop: Select the current range of the signal output.
Make sure the connected device works with the same current
range.
Available ranges: 0–20 mA or 4–20 mA
nd
sample stream
A-96.250.641 / 290312