SFORmat Subsystem
SETHold
an integer from 1 to 12
<pod_num>
integer {0|1|2|3|4|5|6|7|8|9} representing the following setup and
<set_hold_
hold values:
value>
Table 5-2
Setup and hold values
For one clock and one edge
0 = 3.5/0.0 ns
1 = 3.0/0.5 ns
2 = 2.5/1.0 ns
3 = 2.0/1.5 ns
4 = 1.5/2.0 ns
5 = 1.0/2.5 ns
6 = 0.5/3.0 ns
7 = 0.0/3.5 ns
N/A
N/A
Example
OUTPUT XXX;":MACHINE2:SFORMAT:SETHOLD 1,2"
Query
:MACHine{1|2}:SFORMAT:SETHOLD? <pod_num>
The SETHold query returns the current setup and hold settings.
Returned Format
[:MACHine{1|2}:SFORmat:SETHold <pod_num>]
<setup_and_hold_string><NL>
Example
OUTPUT XXX;":MACHINE2:SFORMAT:SETHOLD? 3"
5–14
For one clock and both edges
0 = 4.0/0.0
1 = 3.5/0.5
2 = 3.0/1.0
3 = 2.5/1.5
4 = 2.0/2.0
5 = 1.5/2.5
6 = 1.0/3.0
7 = 0.5/3.5
8 = 0.0/4.0
N/A
Multiple Clocks
0 = 4.5/0.0
1 = 4.0/0.5
2 = 3.5/1.0
3 = 3.0/1.5
4 = 2.5/2.0
5 = 2.0/2.5
6 = 1.5/3.0
7 = 1.0/3.5
8 = 0.5/4.0
9 = 0.0/4.5