HP 16554A Programmer's Manual page 25

State/timing logic analyzers
Table of Contents

Advertisement

Programming the HP 16554A/ HP 16555A/HP 16555D
Mainframe Commands
STOP Command
The STOP command stops the specified module. If the specified module is
configured as part of an intermodule run, STOP will stop all associated
modules.
STARt and STOP are overlapped commands. Overlapped commands allow
execution of subsequent commands while the logic analyzer operations
initiated by the overlapped command are still in progress. For more
information, see *OPC and *WAI commands in Chapter 5 of the
HP 16500/16501A Programmer's Guide.
RMODe Command/query
The RMODe command specifies the run mode (single or repetitive) for a
module. If the selected module is configured for intermodule, the
intermodule run mode will be set by this command. The RMODe query
returns the current setting.
SYSTem:ERRor? Query
The SYSTem:ERRor query returns the oldest error in the error queue. In
order to return all the errors in the error queue, a simple FOR/NEXT loop can
be written to query the queue until all errors are returned. Once all errors
are returned, the query will return zeros.
SYSTem:PRINt Command/query
The SYSTem:PRINt command initiates a print of the screen or listing buffer
over the current printer communication interface. The SYSTem:PRINt query
sends the screen or listing buffer data over the current controller
communication interface.
MMEMory Subsystem
The MMEMory Subsystem provides access to both internal disc drives for
loading and storing configurations.
INTermodule Subsystem
The INTermodule Subsystem commands are used to specify intermodule
arming between multiple modules.