DATA and SETup Commands

SYSTem:DATA

SYSTem:DATA
Command
:SYSTem:DATA <block data>
The SYSTem:DATA command transmits the acquisition memory data from
the controller to the HP 16554A/HP 16555A/HP 16555D logic analyzers.
The block data consists of a variable number of bytes containing information
captured by the acquisition chips. Since no parameter checking is
performed, out-of-range values could cause instrument lockup; therefore,
care should be taken when transferring the data string into the HP 16554A/
HP 16555A/HP 16555D.
The
specifier>
The
represents a digit (ASCII characters "0" through "9"). The value of the eight
D
digits represents the total length of the block (all sections). For example, if
the total length of the block is 14522 bytes, the block length specifier would
be "
Each
The
For the DATA instruction, there is only one
of a data preamble followed by the acquisition data. This section has a
variable number of bytes depending on configuration and amount of acquired
data.
Example
OUTPUT XXX;":SYSTEM:DATA" <block data>
Do not load UNPacked data into the instrument; it may cause the HP 16500 to
lock up. If this happens, cycle power. Only data saved in PACKed mode can be
reloaded back into a logic analyzer.
17–4
parameter can be broken down into a
<block data>
and a variable number of
<block length specifier>
".
#800014522
consists of a
<section>
format varies for each section and may be any length.
<section data>
s.
<section>
always takes the form
and
<section header>
<section data>
, which is composed
<section>
<block length
. Each
#8DDDDDDDD
.