TWAVeform Subsystem

OTIMe

OTIMe
Command
:MACHine{1|2}:TWAVeform:OTIMe <time_value>
The OTIMe command positions the O marker in time when the marker mode
is TIME. If data is not valid, the command performs no action.
real number −2.5 ks to +2.5 ks
<time_value>
Example
OUTPUT XXX; ":MACHINE1:TWAVEFORM:OTIME 30.0E−6"
Query
:MACHine{1|2}:TWAVeform:OTIMe?
The OTIMe query returns the O marker position in time. If data is not valid,
the query returns 9.9E37.
Returned Format
[:MACHine{1|2}:TWAVeform:OTIMe] <time_value><NL>
Example
OUTPUT XXX;":MACHINE1:TWAVEFORM:OTIME?"

OVERlay

Command
:MACHine{1|2}:TWAVeform:OVERlay <module_number>,
<label>,<label>[,<label>]...
The OVERlay command overlays two or more oscilloscope waveforms and
adds the resultant waveform to the current waveforms display. The first
parameter of the command specifies which slot contains the oscilloscope
master card. The next parameters are the labels of the waveforms that are to
be overlaid.
The OVERlay command only works on oscilloscopes. To overlay analyzer
waveforms, use the INSert command with OVERlay option.
13–18