Copy; Data - HP 16554A Programmer's Manual

State/timing logic analyzers
Table of Contents

Advertisement

COMPare Subsystem

COPY

COPY
Command
:MACHine{1|2}:COMPare:COPY
The COPY command copies the current acquired State Listing for the
specified machine into the Compare Listing template. This makes the current
acquisition the reference listing. It does not affect the compare range or
channel mask settings.
Example
OUTPUT XXX;":MACHINE2:COMPARE:COPY"

DATA

Command
:MACHine{1|2}:COMPare:DATA
{<label_name>,<line_num>,<data_pattern>|
<line_num>,<data_pattern>[, <data_pattern>]... }
The DATA command allows you to edit the compare listing image for a given
label and state row. When DATA is sent to an instrument where no compare
image is defined (such as at power-up) all other data in the image is set to
don't cares.
Not specifying the
patterns to more than one label for the given line number. The first pattern
is placed in the leftmost label, with the following patterns being placed in a
left-to-right fashion as seen on the Compare display. Specifying more
patterns than there are labels simply results in the extra patterns being
ignored.
Because don't cares (Xs) are allowed in the data pattern, it must always be
expressed as a string. You may still use different bases, but don't cares
cannot be used in a decimal number.
10–6
parameter allows you to write data
<label_name>

Advertisement