Sem Subsection - Agilent Technologies N9340B Programmer's Manual

Spectrum analyzer
Table of Contents

Advertisement

5
Command Reference

SEM Subsection

72
Set the Average
:MEASure:SEMask:AVERage:COUNt <integer>
:MEASure:SEMask:AVERage:COUNt?
:MEASure:SEMask:AVERage[:STATe] OFF|ON|0|1
:MEASure:SEMask:AVERage[:STATe]?
Specifies the number of measurement averages used when
calculating the measurement result set Avg Number to ON. The
average will be displayed at the end of each sweep.
*RST:
10/Off
Range:
1 to 1000
Key access:
MEAS > SEM > Avg Number On Off
Set the Chan Integ BW
:MEASure:SEMask:BANDwidth|BWIDth:INTegration
<freq>
:MEASure:SEMask:BANDwidth|BWIDth:INTegration?
Specifies the integration bandwidth used in calculating the
power in the main channel.
*RST:
3.84 MHz
Range:
10% to 100% of the setting of Chan Span
Example:
:MEAS:SEM:BAND:INT 4 MHz
:MEAS:SEM:BWID:INT 4 MHz
Key access:
MEAS > SEM > Ref Channel > Chan Integ BW
N9340B Programmer's Guide

Advertisement

Table of Contents
loading

Table of Contents