Agilent Technologies N9340B Programmer's Manual page 60

Spectrum analyzer
Table of Contents

Advertisement

5
Command Reference
Returns a "1" if the last processing is complete. Use this
query when there's a need to monitor the command
execution status, such as a sweep execution.
*OPC and *OPC? are currently effective only when
immediately preceded by eiher the :INITiate:IMMediate
or a :CALibration command.
Reset
*RST
This command presets the instrument to a factory defined
condition that is appropriate for remote programming
operation. *RST is equivalent to performing the two
commands :SYSTem:PRESet and *CLS. This command
always performs a factory preset.
The preset performed by
is always a factory preset. That is, the
N O T E
*RST
same preset performed by
when
:SYSTem:PRESet
:SYS-
is set to
.
Tem:PRESet:TYPE
FACTory
Key access:
Preset
Service Request Enable
*SRE <integer>
*SRE?
This command enables the desired bits of the service request
enable register.
The query returns the value of the register, indicating which
bits are currently enabled. The default value is 255.
Example:
*SRE 16 enables bits 4 in the service request enable register.
Range:
Integer, 0 to 255
54