Agilent Technologies N9340B Programmer's Manual page 70

Spectrum analyzer
Table of Contents

Advertisement

5
Command Reference
64
Trace Y-Axis Reference Level Offset
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel:OFFSet
<ampl>
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel:OFFSet?
This command sets the amplitude level offset for the Y- Axis.
*RST:
0 dB
Range:
–327.60 to +327.60 dB
Key access:
AMPTD > Ref Offset
Vertical Axis Scaling
:DISPlay:WINDow:TRACe:Y[:SCALe]:SPACing LIN-
ear|LOGarithmic
:DISPlay:WINDow:TRACe:Y[:SCALe]:SPACing?
Toggles the vertical graticule divisions between logarithmic
or linear units. The default logarithmic unit is dBm, and the
linear unit is mV.
*RST:
Log
Key access:
AMPTD > Scale Type
Brightness
:DISPlay:MODE:BRIGhtness <integer>
:DISPlay:MODE:BRIGhtness?
This command adjusts the brightness of the current screen
display. Set to 0 to turn off the backlight of the LCD screen.
Range:
0 to 10
Key Access:
SYS > Brightness