N9340B Programmer's Guide
<ampl>,...[<ampl>] (up to five values)
:MEASure:SEMask:OFFSet:LIST:STARt:RCARrier?
:MEASure:SEMask:OFFSet:LIST:STOP:RCARrier:COU-
Ple OFF|ON|0|1,...OFF|ON|0|1 (up to five values)
:MEASure:SEMask:OFFSet:LIST:STOP:RCARrier:COU-
Ple?
Enables you to enter an absolute level limit at Stop Freq for
the currently offset ranging from –150.00 to +50.00 dB with
0.01 dB resolution, and to toggle this function between
Couple and Man. If set to Couple, this is coupled to Rel Start to
make a flat limit line. If set to Man, Abs Start and Rel Stop you
can enter different values to make a sloped limit line.
*RST:
Dependent upon Offset selected, please refer to the User's
Guide for more details.
Range:
–150 dB to 50 dB
Example:
:MEAS:SEM:OFFS:LIST:STOP:RCAR –20 dB
:MEAS:SEM:OFFS:LIST:STOP:RCAR:COUP 0
Key access:
MEAS > Spectrum Emission Mask > Offset/Limits > More > Rel Stop
Set the Fail Mask
:MEASure:SEMask:OFFSet:LIST:TEST ABSo-
lute|AND|OR|RELative,ABSolute|AND|OR|RELa-
tive,ABSolute|AND|OR|RELative,ABSolute|AND|OR|RE
Lative
:MEASure:SEMask:OFFSet[n]:LIST:TEST?
Displays the menu to select one of the following logic keys
for fail conditions between the measurement results and the
test limits: Absolute, Relative, Abs AND Rel, and Abs OR Rel.
Please refer to User's Guide for further descriptions.
*RST:
Dependent upon Offset selected, please refer to the User's
Guide for more details.
Example:
:MEAS:SEM:OFFS:LIST:TEST ABS,REL,ABS AND REL,ABS
OR REL
Command Reference