Agilent Technologies N9340B Programmer's Manual page 79

Spectrum analyzer
Table of Contents

Advertisement

N9340B Programmer's Guide
Set the Channel Res BW
:MEASure:SEMask:BANDwidth|BWIDth[:RESolu-
tion]:AUTO OFF|ON|0|1
:MEASure:SEMask:BANDwidth|BWIDth[:RESolu-
tion]:AUTO?
:MEASure:SEMask:BANDwidth|BWIDth[:RESolution]
<freq>
:MEASure:SEMask:BANDwidth|BWIDth[:RESolution]?
Specifies the resolution bandwidth used in measuring and
the power in the main channel.
*RST:
30 kHz
Range:
10% to 100% of the setting of Chan Span
Example:
MEAS:SEM:BAND 4 MHz
MEAS:SEM:BWID 4 MHz
Key access:
MEAS > SEM > Ref Channel > Res BW
Set Total Power Reference
:MEASure:SEMask:CARRier[:POWer] <ampl>
:MEASure:SEMask:CARRier[:POWer]?
Use this command to set the Total Power Reference value.
The query return a value of total power, with current ampt
unit.
Key access:
MEAS > SEM > Ref Channel > Total Pwr Ref
Set Chan Span
:MEASure:SEMask:FREQuency:SPAN <freq>
:MEASure:SEMask:FREQuency:SPAN?
Specifies the span used in measuring the power in the main
channel.
*RST:
5 MHz
Example:
:MEAS:SEM:FREQ:SPAN 4 MHz
Key access:
MEAS > SEM > Chan Span
Command Reference
5
73

Advertisement

Table of Contents
loading