Sense Subsystem; [:Sense]:Average Subsection - Agilent Technologies N9340B Programmer's Manual

Spectrum analyzer
Table of Contents

Advertisement

SENSe Subsystem

[:SENSe]:AVERage Subsection

N9340B Programmer's Guide
This substem provides the command used to set the
instrument state parameters so that you can measure the
input signal.
Set the Average Count
[:SENSe]:AVERage:TRACe[1]|2|3|4:COUNt <integer>
[:SENSe]:AVERage:TRACe[1]|2|3|4:COUNt?
Specifies the number of measurements that are combined.
*RST:
1
Range:
1 to 1000
Key access:
TRACE > Average
Turn Averaging On/Off
[:SENSe]:AVERage:TRACe[1]|2|3|4[:STATe]
OFF|ON|0|1
[:SENSe]:AVERage:TRACe[1]|2|3|4[:STATe]?
This command toggles averaging off and on. Averaging
combines the value of successive measurements to average
out measurement variations.
*RST:
Off
Key access:
TRACE > Average On Off
Command Reference