5
Command Reference
78
:MEASure:SEMask:OFFSet:LIST:STARt:RCARrier?
Enables you to enter a relative level limit at Start Freq ranging
from –150.00 to +50.00 dB with 0.01 dB resolution.
*RST:
Dependent upon Offset selected, please refer to the User's
Guide for more details.
Range:
–150 dB to 50 dB
Example:
:MEAS:SEM:OFFS:LIST:STAR:RCAR –20 dB
Key access:
MEAS > Spectrum Emission Mask > Offset/Limits > More > Rel Start
Set the Abs Stop
:MEASure:SEMask:OFFSet:LIST:STARt:ABSolute
<ampl>,...[<ampl>] (up to five values)
:MEASure:SEMask:OFFSet:LIST:STARt:ABSolute?
:MEASure:SEMask:OFFSet:LIST:STOP:ABSolute:COU-
Ple OFF|ON|0|1,...OFF|ON|0|1 (up to five values)
:MEASure:SEMask:OFFSet:LIST:STOP:ABSolute:COU-
Ple?
Enables you to enter an absolute level limit at Stop Freq for
the currently offset ranging from –200.00 to +50.00 dBm
with 0.01 dB resolution, and to toggle this function between
Couple and Man. If set to Couple, this is coupled to Abs Start
to make a flat limit line. If set to Man, Abs Start and Abs Stop
you can enter different values to make a sloped limit line.
*RST:
Dependent upon Offset selected, please refer to the User's
Guide for more details.
Range:
–200 dBm to 50 dBm
Example:
:MEAS:SEM:OFFS:LIST:STOP:ABS –20 dBm
:MEAS:SEM:OFFS:LIST:STOP:ABS:COUP 0
Key access:
MEAS > Spectrum Emission Mask > Offset/Limits > More > Abs Stop
Set the Rel Stop
:MEASure:SEMask:OFFSet:LIST:STARt:RCARrier
N9340B Programmer's Guide