5
Command Reference
76
Set the Start Freq
:MEASure:SEMask:OFFSet:LIST:FREQuency:STARt
<freq>,...[<freq>] (up to five values)
:MEASure:SEMask:OFFSet:LIST:FREQuency:STARt?
:MEASure:SEMask:OFFSet:LIST:STATe
OFF|ON|0|1,...OFF|ON|0|1 (up to five values)
:MEASure:SEMask:OFFSet:LIST:STATe?
Specifies the inner limit (frequency closest to the carrier) for
both segments of the specified offset pair. When Start Freq
(Off) is selected, the offset pair is not measured.
Offsets that are turned off, for the currently selected offset
will return a - 999.0 when queried.
*RST:
Dependent upon Offset selected, please refer to the User's
Guide for more details.
Range:
0 Hz to the Stop Freq (for that offset) minus 10 Hz
Example:
:MEAS:SEM:OFFS:LIST:FREQ:STAR 2 MHz
:MEAS:SEM:OFFS:LIST:STAT 1
Key access:
MEAS > Spectrum Emission Mask > Offset/Limits > Start Freq
Set the Stop Freq
:MEASure:SEMask:OFFSet:LIST:FREQuency:STOP
<freq>,...[<freq>] (up to five values)
:MEASure:SEMask:OFFSet:LIST:FREQuency:STOP?
Specifies the inner limit (frequency closest to the carrier) for
both segments of the specified offset pair. When Start Freq
(Off) is selected, the offset pair is not measured.
Offsets that are turned off, for the currently selected offset
will return a - 999.0 when queried.
*RST:
Dependent upon Offset selected, please refer to the User's
Guide for more details.
Range:
0 Hz to the Stop Freq (for that offset) minus 10 Hz
Key access:
MEAS > Spectrum Emission Mask > Offset/Limits > Stop Freq
N9340B Programmer's Guide