What is cyclic pht f02 alarm ?? And management?
1 comments:
Mr. Anderson
March 3, 2025
The Cyclical PHT F02 alarm on the Fresenius Medical Care 4008 E indicates a balancing error. Possible causes include system leakage or, for Diasafe machines, DIP switch array 2, switch 1 on CPU II not being set to "ON". To manage this alarm, check for system leaks and ensure that the DIP switch is correctly set if using a Diasafe machine.
Need help?
Do you have a question about the 4008 E and is the answer not in the manual?
Questions and answers
What is cyclic pht f02 alarm ?? And management?
The Cyclical PHT F02 alarm on the Fresenius Medical Care 4008 E indicates a balancing error. Possible causes include system leakage or, for Diasafe machines, DIP switch array 2, switch 1 on CPU II not being set to "ON". To manage this alarm, check for system leaks and ensure that the DIP switch is correctly set if using a Diasafe machine.