Setting Vapor; Aspirating Secretions - Dräger Julian Operating Instructions Manual

Anesthesia workstation
Hide thumbs Also See for Julian:
Table of Contents

Advertisement

Operation

Setting Vapor

Aspirating Secretions

Setting Vapor
1 Lock the unused Vapor
= push lever fully in the direction of the unused Vapor
(example: left Vapor locked).
2 Press the » O « button and
3 set the handwheel to the required anesthetic agent
concentration.
Aspirating Secretions
Swing the suction system forwards.
1 Set switch to » I «.
2 Seal the secretion probe adapter or kink the suction
hose,
3 Set the appropriate suction pressure for the patient
with the rotary knob »Vac.« and
4 check vacuum on the pressure gauge.
NOTE: Graduation indicators are for approximation only.
WARNING !
Always follow accepted hospital procedures
for handling body fluids.
After aspirating
Rinse the suction hose with the water purifier.
1 Set switch to » 0 «.
56
1
0
D
4
3 1
2
3
0
2
Julian