System Overview
Vaporizers (Optional)
The Dräger Vapor anesthetic agent vaporizers are
used to enrich the fresh gas with a precisely mete-
red quantity of vapor from the liquid anesthetic
agent being used, i.e., Isoflurane, Halothane, Enflu-
rane, Sevoflurane, or Desflurane.
When using a Desflurane vaporizer, it must be con-
nected to mains power.
For complete information, consult the appropriate
Instruction for Use provided with the vaporizer.
Vaporizer Exclusion Systems
The exclusion systems available for the
Fabius plus are described below.
Dräger Vapor Interlock 2 System (Optional)
The Dräger Interlock 2 system is used to ensure
that only one of two vaporizers can be used at a
time. It has a selector lever used to select which
vaporizer is enabled.
Moving the selector lever away from the desired
vaporizer allows that vaporizer to be used and the
other to be locked out of use.
Note that the selector lever is shown in the center
position. This ensures that both vaporizers are in
the locked position. Also, this is the recommended
position for the selector lever when moving the
Fabius plus.
21
Instructions for Use Fabius plus SW 3.n