Vaporizers (Optional)

The Dräger Vapor anesthetic agent vaporizers are
used to enrich the fresh gas with a precisely
metered quantity of vapor from the liquid anesthetic
agent being used, i.e., Isoflurane, Halothane, Enflu-
rane, or Sevoflurane.
WARNING
Do not use any type of Desflurane vaporizer in
the MR environment. In an MR environment
functionality of the Desflurane vaporizer will
be compromised.
CAUTION
Only Vapor 2000 vaporizers can by used on the
Fabius MRI in MRT scanner rooms.
For complete information, consult the appropriate
Instruction for Use provided with the vaporizer.

Vaporizer Exclusion Systems

The exclusion systems available for the
Fabius MRI are described below.
Dräger Vapor Interlock 2 System (Optional)
The Dräger Interlock 2 system is used to ensure
that only one of two vaporizers can be used at a
time. It has a selector lever used to select which
vaporizer is enabled.
Moving the selector lever away from the desired
vaporizer allows that vaporizer to be used and the
other to be locked out of use.
22
Instructions for Use Fabius MRI SW 3.n