Advance Marker Functions - Keysight HSA N9342C Demonstration Manual

Handheld spectrum analyzer
Table of Contents

Advertisement

19 | Keysight | N9342C/N9343C/N9344C HSA Handheld Spectrum Analyzer - Demonstration Guide
Demonstration 9

Advance marker functions

The N934xC HSAs provide powerful marker functions. In addition to normal and delta
marker modes, the instrument marker system provides noise markers, frequency counter
markers, AM/FM tune and listen, and band power measurement. For the AM/FM tune
and listen feature, the AM/FM demodulated audio can be heard with the internal
speaker or the provided headphone. Speaker volume and delay time can be adjusted to
meet your specific needs.
In this demonstration, we will perform an AM/FM tune and listen measurement using an
external AM/FM antenna.
Figure 22. AM/FM-tune and listen marker
Instructions for HSA
Connect an AM/FM antenna (such as 88 MHz to 108 MHz) to HSA RF input port
Set start frequency and stop frequency
Turn on the hi sensitivity mode
Note: If Hi-Sensitive is On, parameters below can be set automatically:
– Reference level -50 dBm
– Attenuator 0 dB
– Preamp On
Enter into FM and listen mode
In order to make sound smooth and clear
Set speaker volume
Set delay time
Keystrokes for HSA
[Freq], {Start Freq}, [88], {MHz}, {Stop Freq}, [108], {MHz}
[Amptd], {More 1 of 2},
{Hi-Sensitive (On)}
[Marker], {Function}, {Demod}, {Demod Setting}, {Demod Type (FM)}, rotate
the knob to move the marker to one of the displayed FM broadcast stations
{Speaker Vol}, [50], {Enter}
{Delay Time}, [1], {s}