Directions for use
Message
Stat mode requested too
soon. Clear to retry.
050609
Unable to determine
NIBP; check connections
and tubing for kinks.
05060A
NIBP not calibrated.
Calibrate the module.
05060B
NIBP not functional.
05060C
NIBP not functional.
05060D
NIBP not functional.
05060E
NIBP not functional.
05060F
NIBP not functional.
050610
NIBP not functional.
050611
NIBP not functional.
050612
NIBP not functional.
050613
NIBP not functional.
050614
NIBP not functional.
050615
NIBP not functional.
050616
NIBP not functional.
050617
NIBP not functional.
050618
Possible cause
NIBP excessive stat mode time.
The time between readings is less
than one minute and the readings
plus time between readings result
in the device taking over 15
minutes to complete the averaging
cycle.
NIBP transducers are not matched The transducer are above 5 mmHg
Factory EEPROM checksum error
on NIBP. Units internal
configuration was corrupted
NIBP command not implemented
NIBP wrong data count
NIBP data range error
NIBP no POST error to clear
NIBP cannot clear this POST error Internal malfunction. If the problem
NIBP command not command type Internal malfunction. If the problem
NIBP communication timeout
NIBP response header wrong
NIBP response checksum wrong
Too much NIBP data was received Internal malfunction. If the problem
NIBP FPROM erase error
NIBP FPROM programming error
Invalid NIBP target pressure
Suggested action
Internal malfunction. If the problem
persists, replace the module.
and the difference in pressure is
greater than 40 mmHg. Check cuff
for pinched or obstructed tubing. If
the problem persists, replace the
module.
Calibrate the NIBP module. If the
problem persists, replace the
module.
Internal malfunction. If the problem
persists, replace the module.
Internal malfunction. If the problem
persists, replace the module.
Internal malfunction. If the problem
persists, replace the module.
Internal malfunction. If the problem
persists, replace the module.
persists, replace the module.
persists, replace the module.
Internal malfunction. If the problem
persists, replace the module.
Internal malfunction. If the problem
persists, replace the module.
Internal malfunction. If the problem
persists, replace the module.
persists, replace the module.
Internal malfunction. If the problem
persists, replace the module.
Internal malfunction. If the problem
persists, replace the module.
Internal malfunction. If the problem
persists, replace the module.
Troubleshooting 93
Alarm priority
Very low
Medium
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low