Welch Allyn Connex Spot Monitor Directions For Use Manual page 120

Hide thumbs Also See for Connex Spot Monitor:
Table of Contents

Advertisement

114 Troubleshooting
Message
Temperature not functional.
3F051A
Temperature not functional.
3F0821
Temperature not functional.
3F0822
Temperature not functional.
3F0824
Temperature not functional.
3F0833
Temperature not functional.
3F0E04
Unable to detect new
temperature. Retry measurement.
Thermometer might be docked
improperly. Check contacts and
connections.
Temperature not functional.
3FFF01
Temperature not functional.
3FFF02
Temperature not functional.
3FFF03
Temperature not functional.
3FFF04
Re-dock Braun. 3FFF05
Possible cause
Suggested action
The reference voltage
Internal malfunction. If the problem
circuit was detected to
persists, replace the module.
be under voltage or
unstable.
The ambient
Verify conditions are less than 104°F or
temperature is too high
40°C. If conditions are valid and the
problem persists, replace the probe. If the
problem still persists, replace the module.
The ambient
Verify conditions are greater than 50°F or
temperature is too low
10°C. If conditions are valid and problem
persists, replace the probe. If the problem
still persists, replace the module.
The battery is over the
Internal malfunction. If the problem
maximum voltage
persists, replace the module.
The sensor not
Internal malfunction. If the problem
functional
persists, replace the module.
Low battery
Recharge the battery. If the problem
persists, check the battery.
No temperature
If a measurement should have been
measurement was
available, retry the measurement. If
available from the
problem the persists, replace the module.
thermometer at the
time it was docked.
Communication failure
The thermometer might be docked
with docked Braun
improperly. Check contacts and
connections. If problem the persists,
replace the module.
Unrecognized WACP
Internal malfunction. If the problem
parameter received
persists, replace the module.
from sensor
Timeout waiting for
Internal malfunction. If the problem
sensor response
persists, replace the module.
Error deserializing
Internal malfunction. If the problem
WACP message
persists, replace the module.
received from sensor
WACP stack send
Internal malfunction. If the problem
message failure
persists, replace the module.
Anti-theft timer expired Re-dock the thermometer after taking a
measurement.
®
Connex
Spot Monitor
Alarm
priority
Very low
Very low
Very low
Very low
Very low
Very low
Information
Information
Very low
Very low
Very low
Very low
Very low