112 Troubleshooting
Message
Temperature not functional.
03CA00
Temperature not functional.
03CB00
Temperature not functional.
03CC00
Temperature not functional.
03CD00
Temperature not functional.
03CE00
Temperature not functional.
03CE01
Direct mode reading timed out
Tissue contact lost.
Temperature module reset.
03D000
Braun 6000 messages
Message
Temperature not functional.
3F0105
Temperature not functional.
3F0201
Temperature not functional.
3F0202
Temperature not functional.
3F0203
Possible cause
Suggested action
Unsupported message
Internal malfunction. If the problem
received from the
persists, replace the module.
Temperature module
Unable to send
Internal malfunction. If the problem
message to the
persists, replace the module.
Temperature module
Temperature module
Internal malfunction. If the problem
communication times
persists, replace the module.
out
Failed to upgrade the
Internal malfunction. If the problem
temperature module
persists, replace the module.
Unable to read PIM file Retry the device update.
Upgrade file directory
Retry the device update
could not be accessed
Direct mode reading
Direction mode reading times out
times out
Lost tissue contact
Check the tissue contact and try the
while attempting to
measurement again.
acquire a temperature
measurement or
acquired measurement
was performed with
limited tissue contact.
Temperature sensor
None
reset unexpectedly
Possible cause
Suggested action
WACP message CRC
Internal malfunction. If the problem
mismatch.
persists, replace the module.
This message is not
Internal malfunction. If the problem
implemented by the
persists, replace the module.
module.
This message is not
Internal malfunction. If the problem
supported by the
persists, replace the module.
module.
The module has run out
Internal malfunction. If the problem
of memory.
persists, replace the module.
®
Connex
Spot Monitor
Alarm priority
Very low
Very low
Very low
Very low
Very low
Very low
Information
Information
Very low
Alarm
priority
Very low
Very low
Very low
Very low

Hide quick links: