Message
Bluetooth not functional.
370001
Bluetooth not functional.
370002
Bluetooth device
connection successful
Bluetooth device
disconnected

APM messages

Message
APM not functional.
1C1001
APM not functional.
1C100B
APM battery is absent
or faulty. 1C100F
The APM is
disconnected. 1C1002
USB cable is
disconnected. 1C1003
Possible cause
The monitor detected a Bluetooth
device that is not functional
The monitor cannot detect a
Bluetooth module
The Bluetooth device connected
The Bluetooth disconnected
Possible cause
The APM is detected as connected
but there is no communication
through the APM serial port
The APM battery is installed, but
does not communicate with the
monitor
The APM battery is not installed
The APM is unplugged from the
monitor while the monitor is
powered on
The APM USB hub is unplugged
from the monitor while the
monitor is powered on
Suggested action
Reboot the monitor. If the problem persists,
replace the Bluetooth radio. If the problem
persists, replace the main PCBA.
Replace the Bluetooth radio. If the problem
persists, replace the main PCBA.
None.
None.
Suggested action
Restart the monitor and the APM. If the
problem is still present, check the cable
connections from the monitor to APM. If the
problem still present, replace the APM. If the
message persists, replace the main PCBA on
the monitor.
Perform diagnostic checks on the monitor. If
the problem is still present, replace the APM
battery. If the problem still present, replace
the APM. If the message persists, replace
the main PCBA on the monitor.
Ensure than an APM battery is installed, and
install one if it is missing. If the problem is
still present, perform diagnostic checks on
the monitor. If the problem is still present,
replace the APM. If the message persists,
replace the main PCBA on the monitor.
Restart the monitor and the APM. If the
problem is still present, check the cable
connections from the monitor to APM. If the
problem still present, replace the APM. If the
message persists, replace the main PCBA on
the monitor.
Restart the monitor and the APM. If the
problem is still present, check the cable
connections from the monitor to APM. If the
problem still present, replace the APM. If the
message persists, replace the main PCBA on
the monitor.
Troubleshooting 127
Alarm
priority
Very low
Very low
Information
Information
Alarm
priority
Very low
Very low
Very low
Very low
Very low