Welch Allyn Connex Spot Monitor Directions For Use Manual page 139

Hide thumbs Also See for Connex Spot Monitor:
Table of Contents

Advertisement

Directions for use
2 Cell battery specifications
Acute care continuous 10 minute cycles - 49 patient cycles - BP, temp, SpO2, no radio,
no scanner (Nellcor)
Acute care continuous 10 minute cycles - 50 patient cycles - BP, temp, SpO2, no radio,
no scanner (Nonin)
Acute care continuous 10 minute cycles - 49 patient cycles - BP, temp, SpO2, no radio,
no scanner (Masimo)
Acute care continuous 10 minute cycles - 41 patient cycles - BP, temp, SpO2, radio,
scanner (Nellcor)
Acute care continuous 10 minute cycles - 41 patient cycles - BP, temp, SpO2, radio,
scanner (Nonin)
Acute care continuous 10 minute cycles - 41 patient cycles - BP, temp, SpO2, radio,
scanner (Masimo)
1
Nellcor is the default for these examples.
Nurse Call specifications
Nurse Call connection specifications
Nurse Call
NIBP specifications
NIBP specifications
Cuff pressure range
Systolic range
Diastolic range
Cuff Inflation Target
Maximum Target Pressure
1
25 V AC or 60 V DC maximum at 1A maximum
Meets or exceeds IEC/ISO 80601-2-30 standards for cuff
pressure range
Adult: 30 to 260 mmHg (StepBP, SureBP)
Pediatric: 30 to 260 mmHg (StepBP, SureBP)
Neonate: 20 to 120 mmHg (StepBP)
Adult: 20 to 220 mmHg (StepBP, SureBP)
Pediatric: 20 to 220 mmHg (StepBP, SureBP)
Neonate: 10 to 110 mmHg (StepBP)
Adult:160 mmHg (StepBP)
Pediatric: 140 mmHg (StepBP)
Neonate: 90 mmHg (StepBP)
Adult: 280 mmHg (StepBP, SureBP)
Pediatric: 280 mmHg (StepBP, SureBP)
Neonate: 130 mmHg (StepBP)
Specifications 133
Hours of use
8.22
8.37
8.29
6.84
6.96
6.90