110 Troubleshooting
Message
Temperature not functional.
30814
Temperature not functional.
30815
Temperature not functional.
30816
Temperature not functional.
30817
Unable to detect new
temperature. Retry measurement.
30818
Unable to detect new
temperature. Retry measurement.
30819
Temperature not functional.
3081A
Temperature not functional.
3081B
Temperature not functional.
3081C
Temperature not functional.
3081D
Replace temperature probe.
3081E
Replace temperature probe.
3081F
Temperature not functional.
30820
Possible cause
Suggested action
The temperature
Internal malfunction. If the problem
module TEMP CONFIG
persists, replace the module.
ACQUIRE FAILURE
The temperature
Internal malfunction. If the problem
module TEMP CONFIG
persists, replace the module.
RELEASE FAILURE
The temperature
Internal malfunction. If the problem
module TEMP CONFIG
persists, replace the module.
INVALID PTR FAILURE
The temperature
Internal malfunction. If the problem
module internal error.
persists, replace the module.
EEPROM not initialized
The temperature
Probe malfunction. Replace probe. If the
module heater
problem persists, replace the module.
indicates on when
turned off.
The temperature
Probe malfunction. Replace probe. If the
module heater
problem persists, replace the module.
indicates off when
turned on.
The temperature
Internal malfunction. If the problem
module HTR_Q is on
persists, replace the module.
and HTRC is off but still
has voltage.
The temperature
Internal malfunction. If the problem
module HTR_Q is tri-
persists, replace the module.
stated with HTRC
enabled and has heater
power.
The temperature
Internal malfunction. If the problem
module turned Q&C on
persists, replace the module.
and the heater voltage
is not high enough.
The temperature
Internal malfunction. If the problem
module heater
persists, replace the module.
hardware failsafe
should have turned off
but did not.
The temperature
Probe malfunction. Replace probe. If the
module probe is above
problem persists, replace the module.
112°F or 43.3°C.
The temperature
Probe malfunction. Replace probe. If the
module has excessive
problem persists, replace the module.
heater energy
The temperature
Internal malfunction. If the problem
module host interface
persists, replace the module.
error
®
Connex
Spot Monitor
Alarm priority
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low