108 Troubleshooting
Message
Ambient temperature out of
range. Clear to retry. 30801
Ambient temperature out of
range. Clear to retry. 30802
Temperature not functional.
30803
Temperature not functional.
30804
Temperature not functional.
30805
Temperature not functional.
30806
Ambient temperature out of
range. Clear to retry. 30807
Replace temperature probe.
30808
Possible cause
Suggested action
detected to be under
voltage or unstable.
The temperature
Verify conditions are greater than 50 °F or
module measurement is
10 °C. If conditions are valid and the
below the allowable
problem persists, replace the probe. If the
temperature values and
problem still persists, replace the module.
beyond the low
ambient or patient
limits.
The temperature
Verify conditions are less than 104 °F or
module measurement is
40 °C. If conditions are valid and the
above the allowable
problem persists, replace the probe. If the
temperature values and
problem still persists, replace the module.
beyond the high
ambient or patient
limits.
The temperature
Internal malfunction. If the problem
module internal
persists, replace the module.
calibration resistor
(RCAL) on the board is
damaged or
contaminated (pulse
too long).
The temperature
Internal malfunction. If the problem
module internal
persists, replace the module.
calibration resistor
(RCAL) on the board is
damaged or
contaminated (pulse
too short).
The temperature
Internal malfunction. If the problem
module internal circuit
persists, replace the module.
validation resistor (PTB)
on the board is
damaged (value is
over).
The temperature
Internal malfunction. If the problem
module internal circuit
persists, replace the module.
validation resistor (PTB)
on the board is
damaged (value is
under).
The temperature
Verify conditions are greater than 50 °F or
module A/D
10 °C. If conditions are valid and the
measurement timed out
problem persists, replace the probe. If the
problem still persists, replace the module.
The temperature
Probe malfunction. Replace the probe. If
module probe was not
the problem persists, replace the module.
characterized/
calibrated
®
Connex
Spot Monitor
Alarm priority
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low