82 Maintenance and service
1. Set the monitor on a flat surface with the screen facing downward to access the
battery cover.
2. Locate the battery cover, indicated by
3. Using a double-slotted screwdriver, loosen the captive screw at the base of the
battery cover, and then remove the cover.
4. Remove the old battery from the battery compartment.
5. Disconnect the battery connector from the battery connection port on the monitor.
6. Insert the battery connector for the new battery into the battery connection port on
the monitor.
7. Insert the new battery into the battery compartment.
8. Replace the battery cover, and then tighten the captive screw at the bottom of the
battery cover.
Note

Replace the APM work surface battery

Before removing the APM work surface battery, power down the monitor and
disconnect the power cord from the mains outlet.
Note
1. Loosen the captive screw on the bottom of the APM work surface that secures the
battery cover.
Do not overtighten the screw.
You do not have to remove the APM work surface from the stand to
remove the APM work surface battery.
.
®