100 Troubleshooting
Message
SpO2 rebooting. 042500 There is an SpO2 SRAM task
SpO2 rebooting. 042600 There is an SpO2 database failure A malfunction is detected. There are
SpO2 rebooting. 042700 There is an invalid SpO2 flash
SpO2 rebooting. 042800 There is a problem with the SpO2
SpO2 rebooting. 042900 There is a problem with the SpO2
Possible cause
queue overrun
memory device
analog ground
digital ground
Suggested action
removed. Second, the board has an
actual hardware failure and recovery
is not possible. It is recommended
that the SpO2 module be replaced,
and if problem is still present replace
the monitor's main board.
A malfunction is detected. There are
two possible causes of these
failures. First, the power provided to
the board is out of specification. In
this case, the failure may be cleared
once the underlying cause is
removed. Second, the board has an
actual hardware failure and recovery
is not possible. It is recommended
that the SpO2 module be replaced,
and if problem is still present replace
the monitor's main board.
two possible causes of these
failures. First, the power provided to
the board is out of specification. In
this case, the failure may be cleared
once the underlying cause is
removed. Second, the board has an
actual hardware failure and recovery
is not possible. It is recommended
that the SpO2 module be replaced,
and if problem is still present replace
the monitor's main board.
A malfunction is detected. There are
two possible causes of these
failures. First, the power provided to
the board is out of specification. In
this case, the failure may be cleared
once the underlying cause is
removed. Second, the board has an
actual hardware failure and recovery
is not possible. It is recommended
that the SpO2 module be replaced,
and if problem is still present replace
the monitor's main board.
A malfunction is detected. There are
two possible causes of these
failures. First, the power provided to
the board is out of specification. In
this case, the failure may be cleared
once the underlying cause is
removed. Second, the board has an
actual hardware failure and recovery
is not possible. It is recommended
that the SpO2 module be replaced,
and if problem is still present replace
the monitor's main board.
A malfunction is detected. There are
two possible causes of these
failures. First, the power provided to
®
Connex
Spot Monitor
Alarm priority
Very low
Very low
Very low
Very low
Very low