70 Patient monitoring
Braun ThermoScan
The Braun ThermoScan Pro 6000 thermometer enables you to transfer an ear
temperature measurement to the monitor.
Read the thermometer manufacturer's directions for use before attempting to configure,
use, troubleshoot, or maintain the thermometer.
Take a temperature at the ear site
®
PRO 6000 thermometer
WARNING Liquids can damage electronics inside the thermometer.
Prevent liquids from spilling on the thermometer. If liquids are spilled on
the thermometer, dry off the thermometer with a clean cloth. Check for
proper operation and accuracy. If liquids possibly entered the thermometer,
remove the thermometer from use until it has been properly dried,
inspected, and tested by qualified service personnel.
CAUTION Probe covers are disposable, nonsterilized, and single-use. The
thermometer is also nonsterilized. Do not autoclave the thermometer and
probe covers. Ensure that probe covers are disposed of according to facility
requirements or local regulations.
CAUTION The thermometer has no user-serviceable parts. If service is
required, call your nearest Welch Allyn Customer Service or Technical
Support facility.
CAUTION Store the thermometer and probe covers in a dry location, free
from dust and contamination and away from direct sunlight. Keep the
ambient temperature at the storage location fairly constant and within the
range of 50°F to 104°F (10°C to 40°C).
WARNING Probe covers are single-use only. Re-use of a probe cover may
result in spread of bacteria and cross-contamination.
WARNING Inaccurate measurement risk. Use only Braun ThermoScan
probe covers with this thermometer.
WARNING Inaccurate measurement risk. Frequently inspect the probe
window and keep it clean, dry, and undamaged. Fingerprints, cerumen,
dust, and other contaminants reduce the transparency of the window and
result in lower temperature measurements. To protect the window, always
keep the thermometer in the accessory dock when the thermometer is not
in use.
CAUTION Inaccurate measurement risk. Before taking a temperature
measurement, make sure that the ear is free from obstructions and excess
cerumen build-up.
CAUTION Inaccurate measurement risk. The following factors can affect
ear temperature measurements for up to 20 minutes:
The patient was lying on his or her ear.
The patient's ear was covered.
The patient was exposed to very hot or very cold temperatures.
The patient was swimming or bathing.
The patient was wearing a hearing aid or an ear plug.
®
Connex
Spot Monitor