Directions for use
Configure pulse rate alarms
You must be in the Intervals profile to configure the pulse rate alarms.
1. Touch the Alarms tab.
2. Touch the Pulse rate vertical tab.
3. Using either the keypad or ▲ or ▼, enter the desired upper and lower alarm limits.
4. Touch the Home tab.
The new alarm settings appear in the Pulse Rate Alarm Limit control button.
SpO2 alarms
SpO2 alarm signal delays
Alarm signal delays are built into the system and are not accessible by clinicians.
The lower range of signal delays is 50–98%. The upper range of signal delays is 52–
100%.
Refer to the SpO2 manufacturer's directions for use for additional information on alarm
signal delays.
Configure SpO2 alarms
1. Verify that you are using the Intervals profile, which contains the Alarms tab.
2. Touch the Alarms tab.
3. Touch the SpO2 vertical tab.
4. Using the keypad or ▲ or ▼, enter the desired upper and lower alarms limits.
5. Touch the Home tab.
The new alarm settings appear in the Alarm Limit control button.

Modifiers and manual parameters

Modifiers enable you to save additional information for the measurements for a specific
patient:
Standard modifiers are the same on all devices
Custom modifiers are specific to a facility or a unit; custom modifiers are set up
during the initial configuration requested by your facility.
Manual parameters are core measurements that you can enter physically on the monitor,
such as height, weight, temperature, and pain.
WARNING Inaccurate measurement risk. Pulse rate measurements
generated through the blood pressure cuff or through SpO2 are subject to
artifact and might not be as accurate as heart rate measurements
generated through ECG or manual palpation.
Patient monitoring 77