94 Troubleshooting
Message
Check cuff inflation
settings.
Tube type does not
match device
configuration.
NIBP not functional.
05FF01
NIBP not functional.
05FF02
NIBP not functional.
05FF03
NIBP not functional.
05FF04
NIBP not functional.
05FF05
NIBP not functional.
05FF06
NIBP not functional.
05FF07
NIBP not functional.
05FF08
NIBP not functional.
05FF09
NIBP not functional.
05FF0A
NIBP not functional.
05FF0B
NIBP not functional.
05FF0C
Device configuration
issue. 05FF0D
NIBP not functional.
05FF0E
NIBP not functional.
05FF0F
Possible cause
Cuff inflation target was
overridden due to max pressure
being too low
Switching to step BP
Unrecognized WACP parameter
received from sensor
Timeout waiting for sensor
response
Error deserializing the WACP
message received from the sensor
WACP stack send message failure Internal malfunction. If the problem
Timeout waiting for asynchronous
sensor message
One or more undetermined
numerics when reading status
indicates OK
Unrecognized sensor reading
status code
Sensor power up failure
WACP rendezvous failure
Application firmware retrieval
error during POST
Upgrade .pim file is corrupt
Configured upgrade firmware
directory cannot be accessed
Configured parameter (NIBP or
SpO2) used in Intervals is missing
The NIBP sensor unexpectedly
reset
The NIBP sensor firmware failed
to upgrade
Suggested action
Change cuff inflation target or
maximum pressure so that the cuff
inflation target is at least 20 mmHg
lower than maximum pressure.
Change tube type to dual lumen or
change the algorithm configuration
to step BP
Internal malfunction. If the problem
persists, replace the module.
Internal malfunction. If the problem
persists, replace the module.
Internal malfunction. If the problem
persists, replace the module.
persists, replace the module.
Internal malfunction. If the problem
persists, replace the module.
Check connections. Restrict patient
movement.
Internal malfunction. If the problem
persists, replace the module.
Internal malfunction. If the problem
persists, replace the module.
Internal malfunction. If the problem
persists, replace the module.
Internal malfunction. If the problem
persists, replace the module.
Internal malfunction. If the problem
persists, replace the module.
Internal malfunction. If the problem
persists, replace the module.
Use configured parameters for
Intervals
Clear the error and try again
Internal malfunction. If the problem
persists, replace the module.
®
Connex
Spot Monitor
Alarm priority
Information
Information
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low