Directions for use
Cleaning agent
Accel INTERVention
Accel TB
CaviWipes
Clinell® Universal Wipes
Oxiver TB
Sani-Cloth
Super Sani-Cloth
70 percent isopropyl alcohol solution
Section 2. Not approved for all Connex Spot Monitor components
Note
Cleaning agent
Cleancide
Clorox HealthCare Bleach Germicidal Cleaner
Sani-Cloth
Super HDQ
Tuffie5 Cleaning Wipes
Virex II (256)
10 percent bleach solution
Remove liquid spills from the monitor
Liquids can damage electronics inside the monitor. Follow these steps if liquids spill on
the monitor.
1. Power down the monitor.
2. Disconnect the power cord from the mains outlet and the power source.
3. Remove battery pack from the monitor.
4. Dry excess liquid from the monitor.
5. Reinstall battery pack.
6. Reconnect the power cord.
®
Plus
®
The following cleaning agents are NOT approved for cleaning Connex Spot
Monitors equipped with the Braun ThermoScan PRO 6000
®
Bleach
®
L10
Maintenance and service 85
Additional information
Applied to a clean cloth
Additional information
Not approved for cleaning the display
Dilution rate of ½ oz per gallon of water (1:256) applied to a
clean cloth
Dilution rate of ½ oz per gallon of water (1:256) applied to a
clean cloth
(.5% - 1% sodium hypochlorite) applied to a clean cloth