Welch Allyn Connex Spot Monitor Directions For Use Manual page 103

Hide thumbs Also See for Connex Spot Monitor:
Table of Contents

Advertisement

Directions for use
Message
Replace the SpO2
sensor. 040f00
Replace the SpO2
sensor. Replace the
SpO2 cable. 041000
Sensor not connected.
Clear to retry. 041100
The sensor has expired.
Replace the SpO2
sensor. 041200
Incompatible sensor.
Replace the SpO2
sensor. 041300
Incompatible sensor.
Replace the SpO2
sensor. 041400
Replace the SpO2
sensor. 041500
Possible cause
The SpO2 sensor is defective
There was a fault with the SpO2
sensor and cable.
An adhesive SpO2 sensor is not
connected
The adhesive SpO2 sensor is
expired
The adhesive SpO2 sensor is
incompatible
The adhesive SpO2 sensor is not
recognized
The adhesive SpO2 sensor is
defective
Suggested action
Replace SpO2 sensor. If the problem
is still present, replace the SpO2
cable. If the problem is still present,
verify the module functionality by
replacing the sensor with an
applicable SpO2 tester. If the
message persists, replace the
module.
Check the sensor and cable
connection. If the problem is still
present, replace the SpO2 sensor. If
the problem is still present, replace
the cable. If the problem is still
present, verify the module
functionality by replacing the sensor
with an applicable SpO2 tester. If the
message persists replace the
module.
Connect SpO2 sensor. If the problem
is still present, replace the SpO2
cable. If the problem is still present,
verify the module functionality by
replacing the sensor with an
applicable SpO2 tester. If the
message persists, replace the
module.
Replace SpO2 sensor. If the problem
is still present, replace the SpO2
cable. If the problem is still present,
verify the module functionality by
replacing the sensor with an
applicable SpO2 tester. If the
message persists, replace the
module.
Replace SpO2 sensor. If the problem
is still present, replace the SpO2
cable. If the problem is still present,
verify the module functionality by
replacing the sensor with an
applicable SpO2 tester. If the
message persists, replace the
module.
Replace SpO2 sensor. If the problem
is still present, replace the SpO2
cable. If the problem is still present,
verify the module functionality by
replacing the sensor with an
applicable SpO2 tester. If the
message persists, replace the
module.
Replace SpO2 sensor. If the problem
is still present, replace the SpO2
cable. If the problem is still present,
verify the module functionality by
replacing the sensor with an
Troubleshooting 97
Alarm priority
Very low
Very low
Very low
Very low
Very low
Very low
Very low